Digital DC-Reconstruction of AC-Coupled Electrophysiological Signals with a Single Inverting Filter

Research output: Contribution to journalJournal articlepeer-review

  • Roger Abächerli
  • Isaksen, Jonas L.
  • Ramun Schmid
  • Remo Leber
  • Hans-Jakob Schmid
  • Gianluca Generali
Since the introduction of digital electrocardiographs, high-pass filters have been necessary for successful analog-to-digital conversion with a reasonable amplitude resolution. On the other hand, such high-pass filters may distort the diagnostically significant ST-segment of the ECG, which can result in a misleading diagnosis. We present an inverting filter that successfully undoes the effects of a 0.05 Hz single pole high-pass filter. The inverting filter has been tested on more than 1600 clinical ECGs with one-minute durations and produces a negligible mean RMS-error of 3.1*10−8 LSB. Alternative, less strong inverting filters have also been tested, as have different applications of the filters with respect to rounding of the signals after filtering. A design scheme for the alternative inverting filters has been suggested, based on the maximum strength of the filter. With the use of the suggested filters, it is possible to recover the original DC-coupled ECGs from AC-coupled ECGs, at least when a 0.05 Hz first order digital single pole high-pass filter is used for the AC-coupling.
Original languageEnglish
Article numbere0150207
JournalP L o S One
Volume11
Issue number3
Number of pages14
ISSN1932-6203
DOIs
Publication statusPublished - 3 Mar 2016