The Combination of In vivo (124)I-PET and CT Small Animal Imaging for Evaluation of Thyroid Physiology and Dosimetry

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningfagfællebedømt

OBJECTIVE: A thyroid rat model combining functional and anatomical information would be of great benefit for better modeling of thyroid physiology and for absorbed dose calculations. Our aim was to show that (124)I-PET and CT small animal imaging are useful as a combined model for studying thyroid physiology and dose calculation.

METHODS: Seven rats were subjects for multiple thyroid (124)I-imaging and CT-scans. S-values [mGy/MBqs] for different thyroid sizes were simulated. A phantom with spheres was designed for validation of performances of the small animal PET and CT imaging systems.

RESULTS: Small animal image-based measurements of the activity amount and the volumes of the spheres with a priori known volumes showed a good agreement with their corresponding actual volumes. The CT scans of the rats showed thyroid volumes from 34-70 mL.

CONCLUSIONS: The wide span in volumes of thyroid glands indicates the importance of using an accurate volume-measuring technique such as the small animal CT. The small animal PET system was on the other hand able to accurately estimate the activity concentration in the thyroid volumes. We conclude that the combination of the PET and CT image information is essential for quantitative thyroid imaging and accurate thyroid absorbed dose calculation.

OriginalsprogEngelsk
TidsskriftDiagnostics
Vol/bind2
Udgave nummer2
Sider (fra-til)10-22
Antal sider13
ISSN2075-4418
DOI
StatusUdgivet - 2012

ID: 156848956